VIP标识 上网做生意,首选VIP会员| 设为首页| 加入桌面|| 手机版| 无图版| RSS订阅

生物安全柜工作原理示意图

放大字体缩小字体发布日期:2007-07-27
核心提示:Class II Type A1 BSC Characteristics Exhaust approximately 30% of total air handled and recirculate 70% in cabinet Minimum inflow air velocity is 75 fpm (0.38 m/s) through the front opening May exhaust HEPA filtered air back into the room or through

      Class II Type A1 BSC Characteristics
      Exhaust approximately 30% of total air handled and recirculate 70% in cabinet

      Minimum inflow air velocity is 75 fpm (0.38 m/s) through the front opening

      May exhaust HEPA filtered air back into the room or through a canopy (to the outdoors)

      Audiovisual alarm required when connected to a building exhaust system

      May have contaminated ducts and plenums under positive pressure to the room
      Type A1 cabinets are not suitable for work with volatile toxic chemical and volatile radionuclides.

      关键词: 生物安全柜


      推荐图文
      推荐食品图库
      点击排行
      Processed in 0.134 second(s), 17 queries, Memory 0.88 M
      Baidu
      map