生物安全柜工作原理示意图

放大字体缩小字体发布日期:2007-07-27
核心提示:Class II Type A1 BSC Characteristics Exhaust approximately 30% of total air handled and recirculate 70% in cabinet Minimum inflow air velocity is 75 fpm (0.38 m/s) through the front opening May exhaust HEPA filtered air back into the room or through

      Class II Type A1 BSC Characteristics
      Exhaust approximately 30% of total air handled and recirculate 70% in cabinet

      Minimum inflow air velocity is 75 fpm (0.38 m/s) through the front opening

      May exhaust HEPA filtered air back into the room or through a canopy (to the outdoors)

      Audiovisual alarm required when connected to a building exhaust system

      May have contaminated ducts and plenums under positive pressure to the room
      Type A1 cabinets are not suitable for work with volatile toxic chemical and volatile radionuclides.